Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.
Ultrasonic level measurement system for liquids. Measuring range approx. 9 m (DU 100) / 15 m (DU 101). With temp. sensor for compensation of sound velocity. A measuring system (package) consists of the FMU 100 plug-in card in Monorack housing 7 HP with power pack and the sensor DU 100 or 101. According to data sheet supplied.
Level instrument, continuous.
Measurement ultrasonic, contactless.
Membrane/sensor enclosure PP-GF.
Max meas. range under ideal condition:
Liquid 10m/32ft, solid 5m/16ft.
Blocking distance 50cm/1.6ft.
Max operat. temp.-20...80oC/-4...176oF.
max operating pressure 2bar (30psi).
up to -40oC/-40oF with reduced range.
Integrated sound velocity compensation.
Self cleaning effect of sensor membrane.
Protection class IP68 NEMA6P.
:: Weather resistance.
:: Overflow safe.
Sensor cable up to 300m connectable.
Only available for spare part delivery!
Level instrument, continuous.
Measurement ultrasonic, contactless.
Membrane/sensor enclosure PP-GF.
Max meas. range under ideal condition:
Liquid 20m/65ft, solid 10m/32ft.
Blocking distance 80cm/2.6ft.
Max operat. temp.-20...80oC/-4...176oF.
max operating press. 2bar (30psi) abs.
up to -40oC/-40oF with reduced range.
Integrated sound velocity compensation.
Self cleaning effect of sensor membrane.
Protection class IP68 NEMA6P.
:: Weather resistance.
:: Overflow safe.
Sensor cable up to 300m connectable.
Only available for spare part delivery!