Measured variables
Transmittance, opacity, relative opacity, extinction, dust concentration
-
Process temperature
–40 °C ... +600 °C
-
Measuring range
Transmittance :100 ... 50 % / 100 ... 0 %
Opacity: 0 ... + 50 % / 0 ... 100 %
Relative opacity: 0 ... + 50 % / 0 ... 100 %
Extinction: 0 ... + 0.3 / 0 ... 1
Dust concentration: 0 ... + 200 mg/m³ / 0 ... 10,000 mg/m³
The measurement depends on measuring distance and dust properties
Anvendelsesområde
The DUSTHUNTER T50 is an economic measuring device for dust at medium to high concentrations in gases above the dew point. Transmittance is the basic measuring value. Opacity and extinction can be calculated and displayed as well as the dust concentration after gravimetric calibration.
Emission monitoring in heating systems
Monitoring of dust concentrations upstream of filter systems
Dust concentration measurements in cement plant
Device properties:
For medium to high dust concentrations
Automated thorough check of zero and reference point
For low to medium measuring distances
Fordele
Easyinstallation, commissioning, and operation
Measurement not dependent on gas velocity, humidity, or particle load
Self-monitoring for low maintenance
Xpert
-
Specialiserede produkter
- Designet til krævende applikationer
Teknisk ydeevne
Enkelhed
Variabel
-
Standardprodukter
- Pålidelige og robuste med lav vedligeholdelse
Teknisk ydeevne
Enkelhed
-
Specialiserede produkter
- Designet til krævende applikationer
Teknisk ydeevne
Enkelhed
Variabel
FLEX valg | Teknisk ydeevne | Enkelhed |
---|---|---|
Fundamental Få dækket dine basale målebehov |
Teknisk ydeevne
|
Enkelhed
|
Lean Administrer nemt dine kerneprocesser |
Teknisk ydeevne
|
Enkelhed
|
Extended Optimer dine processer med innovative teknologier |
Teknisk ydeevne
|
Enkelhed
|
Xpert Til de mest udfordrende applikationer |
Teknisk ydeevne
|
Enkelhed
Variabel |
-
-
Kompetencebrochure (CP)
Emission Monitoring Solutions – broad instrument portfolio and longstanding experience
Best in class dust measurement technology with the scattered light and
transmittance dust measuring devices from Endress+HauserSprog:
Engelsk
Version:
01/01/2025
Filstørrelse:
3 MB
Filnavn:
8030014_CP_Dust_Measuring_Devices_en_V3-0_2025.01.pdf
-
-
-
Kompetencebrochure (CP)
Emission Monitoring Solutions – broad instrument portfolio and longstanding experience
An overview brochure about emission monitoring, their measuring
principles and technologies, global requirements and services, a
selection guide and the measuring devices at a glanceSprog:
Engelsk
Version:
01/01/2025
Filstørrelse:
4.5 MB
Filnavn:
8029975_CP_Emission_Monitoring_Solutions_en_V3-0_2025.01_EHS.pdf
-
-
-
Verificerings-DTM
DUSTHUNTER T50 SDD file
SOPAS device description file for DUSTHUNTER T50
Sprog:
Language neutral
Version:
05/12/2024
Filstørrelse:
358 KB
Filnavn:
DUSTHUNTER T50 SDD file
-